Document
Moisture-Analysis-Medical-Devices-Plastic-Application-Note-en-L
Contenu du document
The Importance of Accurately Measuring
the Moisture Content of Plastic Resin
in the Molding Process of Medical Devices
King Snyder017%, Dennis Acord017%, Evelyn Marschall2??5
1. Sales Specialist, Moisture Products, Sartorius USA,
2. Product Manager, Moisture Analysis, Sartorius Lab Instruments, Germany
Abstract
Moisture content is an important variable that must be monitored for and controlled during the production of plastic medical
device parts. ASTM standard D6869 is the benchmark for measuring the moisture content of plastic resin, and stipulates
the use of Karl Fischer (KF) titration as the applicable standard method. However, a KF titrator is expensive, and testing for
moisture requires the use of toxic chemicals so that it can only be used by a trained chemist. A loss-on-drying (LOD) moisture
analyzer can be an alternate method for testing moisture in plastic resin, but must be correlated to the KF titration as the
measurement principles of both differ. Herein, we show that the Sartorius Mark 3 High Performance Moisture Analyzer cor-
relates well with KF titration standards for a number of plastic resins commonly used in medical device manufacture.
The Mark 3 is a recognized ASTM 6980 standard method for testing the moisture content of plastic resin that is a good
return on investment for medical device manufacturers because it is easy to use, has low maintenance costs, reduces the risk
of rejected parts and customer dissatisfaction, and reduces energy consumption. October, 2019
Keywords:
Omnimark, Mark 3, Moisture analyzer for medical
devices, plastic moisture analysis, ASTM Moisture Plastic
resins standard, Moisture testing plastic resin
Find out more: www.sartorius.com
Application Note
2
Introduction
One important variable to monitor and control for during
the production of plastic medical device parts is moisture
content. Resin manufacturers provide material specification
sheets detailing the maximum moisture specification, or
the maximum allowable ratio of water in the resin before
molding or extrusion. Depending on the resin, maximum
moisture specification can vary from 0.20 % to as low as
0.005 % moisture.
The manufacturer’s data sheet also provides general
guidelines on drying temperature and time for the resin
before molding. For instance, a typical drying specification
is drying the resin at 180 °F for 4 hours before molding.
However, these general recommendations do not take into
consideration the environment where the resin is stored,
or the efficiency of the drying system being used. When
following the recommendations for drying temperature
and time, the result is likely to be satisfactory – if the resin
is within anticipated moisture levels before loading into
the dryer.
If the resin is wetter than anticipated, however, the recom-
mended drying time may not be enough to drive off the
moisture required to meet the maximum moisture speci-
fication. Molding wet resin can result in splay marks, streak-
ing, fogging, burrs, and poor mechanical properties. Many
medical device parts, such as syringes, medical tubing,
and even eyeglass lenses, must be totally transparent and
without visual defects. Other medical parts are molded to
very precise specifications, and excess moisture can affect
the flow properties of the resin, causing imperfections that
make the final part useless for its intended purpose. Over-drying a resin, on the other hand, can actually be
more problematic than under-drying. An over-dried part
may look perfect, yet, over-drying may burn off technically
important plasticizers and additives that are critical for the
durability, elasticity, and corrosion resistance of the part.
Accurate measurement of the moisture content
in plastic resin
ASTM standard D6869 is the benchmark for measuring
the moisture content of plastic resin, and stipulates the use
of Karl Fischer (KF) titration as the applicable standard
method. However, a KF titrator is both expensive to purchase
and to maintain. Also, testing moisture with a KF titrator
requires the use of toxic chemicals, so the test can only be
Generic Specific GradeKF TitrationMark 3 ResultsTest Times (min.)Std. Dev.
Polypropylene RTP 199 X 991670.014 %0.01 %4.0 0.001
Polystyrene Styron 4780.021 %0.021 %5.0 0.001
Polycarbonate Lexan 5000.015 %0.016 %4.0 0.002
Peek Victrex 381G 0.015 %0.015 %4.3 0.002
ABS Lustran LK 2790.061 %0.060 %10 0.001
Polysulfone Udel P-17000.020 %0.019 %5.0 0.001
Polyethylene Borstar HE3490-LS0.017 %0.016 %5.0 0.001
Polyamide Durethan BG30X0.071 %0.072 %5.0 0.004
PET Rynite 5300.020 %0.019 %3.5 0.003
Polyvinyl Choride Apex 910-R40.033 %0.032 %5.0 0.002
TPC Hytrel 82380.040 %0.040 %6.3 0.001
PBT Valox 3250.010 %0.010 %4.0 0.001
Table 1: Correlation results between Karl Fischer coulometric titration and the Mark 3 moisture analyzer.
3
Conclusions
Manufacturers of medical parts recognize the necessity of
accurately monitoring the moisture content of their plastic
resin throughout the manufacturing process. The high cost
of resins used in making these parts, the stringent specifi-
cation these parts must meet, and the potential liability
should a part fail, all make the purchase of an LOD moisture
analyzer a good investment. The Mark 3 High Performance
Moisture Analyzer is a good choice for such manufacturers.
The Mark 3 can be programmed with methods for all the
resins manufacturers commonly process, and is recognized
as an ASTM 6980 standard method for testing the moisture
content of plastic resin. The Mark 3 is easy to use and has
low maintenance costs. In addition, the cost savings associ-
ated with reducing the risk of rejected parts and customer
dissatisfaction as well as energy reductions by cutting dry-
ing time makes the purchase of a Mark 3 High Performance
Moisture Analyzer a good return on investment.
For more information on the Mark 3, including a quote,
please visit www.sartorius.com
performed by a trained chemist, using appropriate safety
precautions. Therefore, most manufacturers prefer to use
alternative methods for moisture testing.
Fortunately, ASTM 6980 states that a loss-on-drying (LOD)
moisture analyzer can be an alternate method for testing
moisture in plastic resin. LOD moisture analyzers measure
weight loss to determine the moisture content of a sample.
Because they cannot distinguish between weight loss from
water and weight loss from other volatiles being driven off
by heating the sample, ASTM 6980 requires that test meth-
ods on an LOD analyzer are developed to correlate to the
ASTM 6980 KF titration standard.
One such LOD analyzer is the Sartorius Mark 3 High Perfor-
mance Moisture Analyzer, which is recognized as an ASTM
6980 standard method for testing the moisture content of
plastic resin. We undertook this study to demonstrate the
calibration of the Mark 3 to the KF titration standard for a
number of different resins.
Methods & Results
Our Sartorius Moisture Application labs, based in the US
(Bohemia, New York) and Göttingen (Germany) offer pro-
gram development for all kinds of plastic resins, including
correlation to the reference method (usually KF titration).
To make sure that the final program provides traceable and
reproducible results for samples of different moisture con-
tent, dry and wet samples are used during program devel-
opment to make sure the resulting program is sufficient in
any case. Specific programs are developed for each resin
to guarantee both a correlation between KF titration and
the Mark 3 and to ensure ASTM 6980 requirements of are
met. To date, Sartorius has developed programs for over
7,000 resins.
The table below lists some commonly used plastic resins
in the molding and extruding of medical parts, along with
data showing the correlation between the Mark 3 and KF
titration standard. Of particular note, the standard deviation
between the moisture content as determined by the KF
titration standard and the Mark 3 never reaches statistical
significance, indicating that the Mark 3 gives comparable
results to the KF titration standard.
For further contacts, visit
www.sartorius.com
Specifications subject to change without notice.
Copyright Sartorius Lab Instruments GmbH & Co. KG.
Status: 09 | 2020
Germany
Sartorius Lab Instruments GmbH & Co. KG
Otto-Brenner-Strasse 20
37079 Goettingen
Phone +49 551 308 0 USA
Sartorius Corporation
565 Johnson Avenue
Bohemia, NY 11716
Phone +1 631 254 4249
Toll-free +1 800 635 2906
Entreprise(s) concernée(s) :
Date d'upload du document :
vendredi 17 février 2023
Document protégé
Document uniquement accessible aux visiteurs connectés
Pas encore de compte ?
Inscrivez-vous
Déjà un compte ?
Connectez-vous